General Information


1.1 Specifications

CURRENT (mA)

MILLIAMP SECONDS (mAs)

Range Resolution

Input Impedance*

Range Resolution

Threshhold

Current

Input

Impedance*

20 mA

0.01 mA

100 Ohms

200 mA

0.1 mA

10 Ohms

200 mAs

0.1 mAs

3.2 mA

10 Ohms

2000 mA

1 mA

1 Ohm

2000 mAs

1 mAs

32 mA

1 Ohm

SIGNAL INPUT LIMITS

Function

Range Maximum Input Limit

Function

Range

Maximum Input Limit

mA

OFF Input shorted –2.0 A max

(fuse protected)

mAs

20

Input shorted –2.0 A max
(fuse protected)

20 250 mA for 30 sec*

200

1 A for 30 sec*

200 1 A for 30 sec*

2000

2.0 A max (fuse protected)

2000 2.0 A max (fuse protected)

* Limit set by power dissipation rating of shunt resistors. Maximum common mode voltage is 500 V.

GENERAL

Meter Accuracy: ±1% of rdg + 2d on all ranges Input

Jacks: touch - protected 4 mm banana

Battery Type: Eveready type 522 or equivalent

Fuse: 5 x 20 mm type F2AH250V fuse

Battery Life: Approximately 200 hours

Display: 3-1/2 digit LCD; low battery annunciator,
0.50 in high digits

Environmental:

Operating Temperature: 5°C to 35°C

Storage Temperature: -20°C to 50°C

Humidity: 0 to 80%

Test Leads: 91.4 cm (36 in.) safety set with finger
guards on probes and shrouded banana plugs

Dimensions and Weight:

15 cm wide x 5.0 cm deep

x 8.75 cm high (6 in x 2 in x 3-1/2 in);

0.35 kg (0.78 lb)

* Specifications subject to change without notice.