Instrument Features

2

2-3

INPUT JACKS

Touch-protected 4mm laboratory banana jacks for the use of safety laboratory leads and all spring-loaded
4mm banana plugs.

FUSE ASSEMBLY

Allows field replacement of the input current protection fuse with a small flat-blade screwdriver.

DISPLAY

3-1/2 digit liquid crystal display indicates the measured value of the signal applied to the input in mA or
mAs.

RESET SWITCH

Resets the display to zero after a mAs measurement has been made and the display reading has been
recorded.

MODE SWITCH

Selects either AC or DC measurement mode. The position of this switch depends upon the type of
current signal from the x-ray generator to be measured.

FUNCTION SWITCH

Turns the Model 35035 mA/mAs Meter power ON and selects the desired measurement function (mA or
mAs).

RANGE SWITCH

Selects the full-scale measurement range of the Model 35035 (20, 200, or 2000). All three ranges
operate in the mA function. Only the 200 and 2000 ranges operate in the mAs function.

DISPLAY ZERO ADJUST

Trimmer potentiometer (R33) provides adjustment for a zero display reading with no signal applied to the
input jacks.

MA GAIN ADJUST

Trimmer potentiometer (R10) provides calibration adjustment of the mA function.

MAS GAIN ADJUST

Trimmer potentiometer (R7) provides calibration adjustment of the mAs function.