Maintenance, Calibration, and Troubleshooting

Commercial Detectors

4

4-5

Table 4-2.

Recommended Test Equipment

Feature Specification

Pulse Generator (Square Wave)

Pulse

± 0.1 to 5 V variable (into a 50 ohm load)

Rise Time

≤ 30 ns

Pulse Width

1.0 microsecond (variable)

Output Impedance

50 ohms

Maximum Repetition Rate

2 MHz

Digital Voltmeter

Ranges

± 0 to 100 mV up to ± 0 to 1000 V (20%
overrange)

Accuracy

± 0.1% of input +1 digit

Input Impedance

≥ 10 megohms

Oscilloscope (Dual Trace)

Bandwidth

10 MHz @ 10 mV/div

Fastest Sweep Time

< 10 Microseconds/div

Electrostatic Voltmeter

Range

0 to 2000 VDC

Precision

± 1%

Input Impedance

10

12

ohms

Counter/Scaler

Maximum Count Rate

10

5

cps

Minimum Pulse Width for Counting

40 ns

Preset Counting Time

0.1,1.0, and 10 minutes

4.7.2 Dynode Test Measurement


Preamplifier input is easily damaged. Ground pin 3
of Z1 before applying power and measuring
photomultiplier tube anode or dynode voltage.

1. Using an electrostatic voltmeter, measure the total DC resistance of the dynode string. The value

should be 9.67 megohms ± 15%. If no reading is obtained, check dynode resistor interstage
connections.

2. Connect the cable to the readout.

3. Turn on channel power and apply 900 V to the detector. Measure the anode resistor voltage (R13)

using an electrostatic voltmeter. The value should be 900 V.

4. Remove channel power.

4.7.3 Preamplifier Checkout

High voltage must be removed for this test.

1. Remove the grounding jumper from pin 3 of Z1.

2. Connect an oscilloscope to the input circuitry of the readout (50 ohms terminated).