4 functional description – Fluke Biomedical 943-25 User Manual

Page 6

Advertising
background image

Victoreen 943-25, 943-26
Operators Manual

1-2

Response Ratio:

1:1

943-25, 943-26, 943-25T, 943-26T

10:1

943-25A, 943-26A, 943-25TA, 943-26TA

100:1

943-25B, 943-26B, 943-25TB, 943-26TB

End Window Material

Mylar

943-25, 943-25A, 943-25B, 943-26, 943-26A, 943-26B

Titanium 943-25T, 943-25TA, 943-25TB, 943-26T, 943-26TA, 943-26TB

Preamplifier

Maximum Pressure

30 psi

(Face Exposure)

Rise Time

< 60 ns

Input Impedance

> 50 kilohms

Output Coupling

AC

Voltage Gain

6 V/V

Configuration

Voltage sensitive

Output Polarity

Negative

Output Impedance

50 ohms

Maximum Cable Length

1500 ft

Maximum Pulse Amplitude @ -6 V
Maximum Cable Length

Dead Time

Approximately 10 us @ 8 ft cable length

Maximum Count Rate

10

7


The beta detectors have been assembled with parts selected for the reliability required in a nuclear
application. Any unauthorized repairs made to the detectors utilized in nuclear applications will void the
safety-related rating. Safety-related detectors must be returned to Fluke Biomedical, Radiation
Management for authorized, qualified (ANSI 45.2.6, Skill Level II) service.

1.4 Functional Description

The Victoreen beta scintillation detector is shown in block diagram form in Figure 1-1. The detector
consists of a scintillation disc, a photomultiplier tube, a preamplifier assembly, and various interconnecting
cables connecting the detector, readout, and the detector power supply that is located at the ratemeter or
scaler module.













Figure 1-1. Block Diagram, 943 Series Beta Detector