Victoreen 943-25, 943-26
Operators Manual

4-6

3. Turn on channel power.

4. Using a pulse generator, inject negative pulses of -0.50 V amplitude, one microsecond pulse

duration, at a frequency of 1 kHz to the node of R13 and C6 (+) and gnd (-).

5. Output pulses should be -1.6 V ± 20% amplitude. If pulses are not present, check Z1 and
associated circuitry.


If the preamplifier and dynode tests are positive, the detector malfunction is probably in the
photomultiplier tube. Remove all power, replace the photomultiplier tube and reassemble the detector
according to the procedure in Section 4.1.2.

4.8 Replacement Parts

Drawing Number

Description

943-25

Beta Scintillation Detector, Mylar End Window,

943-26

1:1 Response Ratio

943-25A

Beta Scintillation Detector, Mylar End Window,

943-26A

10:1 Response Ratio

943-25B

Beta Scintillation Detector, Mylar End Window,

943-26B 100:1

Response

Ratio

943-25T

Beta Scintillation Detector, Titanium End Window,

943-26T

1:1 Response Ratio

943-25TA

Beta Scintillation Detector, Titanium End Window,

943-26TB

10:1 Response Ratio

943-25TB

Beta Scintillation Detector, Titanium End Window,

943-26TB 100:1

Response

Ratio



Refer to the individual assembly drawings for a
breakdown of individual components.

NOTE