Nuclear Associates 30-471
Operators Manual

2-2

2.2 Photon Detectors

These detectors are used in the measurement of radiation in the range of 1 MV to 25 MV. They are
available with a negative or positive output. They have been color coded for ease of identification.
Following is a list of the detectors, polarity, color codes, and range.

Model Number

Range

Polarity

Color

30-471

1-4 MV Photon

Positive

Blue

30-471-8000

1-4 MV Photon

Negative

Blue

30-472 5-11

MV

Photon

Positive

Yellow

30-472-8000 5-11

MV

Photon

Negative Yellow

30-473

12-17 MV Photon

Positive

Red

30-473-8000

12-17 MV Photon

Negative

Red

30-474

18-25 MV Photon

Positive

Green

30-474-8000

18-25 MV Photon

Negative

Green

30-475

5-25 MeV Electron

Positive

Gray

30-475-8000

5-25 MeV Electron

Negative

Gray


Build-up shields are utilized in their design to provide equilibrium and filtering of low energy radiation.
These build-up shields also increase the detector's stability along with increasing the sensitivity in their
respective ranges. The output of a typical detector is 1.2 nano coulombs/RAD (1.2 nC/cGy).

The semiconducting material in these diodes is planar and the diodes as a whole is hemispherical in
shape, thus they have a preferred direction. Following are typical angular dependence of these photon
detectors.

Angular Dependence

1-4 MV Diodes

The following figures show the results for incident angles rotation along the diode cable axis and along
the axis perpendicular to diode cable axis. For angle deflections less than 50 degrees from the diodes
preferred direction, 180 degrees, the diode output changed by less than 1%.

Gantry angle (degree with 180 being vertical)

Diode (type 471 [1–4 MV rated]) output vs. incident angle;

symmetric setup in 6 MV photons

Diode out

p

ut (n

ormalized

to

180°)