Nuclear Associates 30-471
Operators Manual

2-4
























12-17 MV Diodes

As seen in the following figures, the results are good for angle variations in the plane perpendicular to the
direction of the cable. Again, there is a small (less than 1%) variation extending to ± 30-35 degrees off
center. Along the plane that includes the cable there is a 3% - 4% variation over the same angular range.



Gantry angle (degree with 180 being vertical)

Diode (type 472 [5–11 MV rated]) output vs. incident angle;

asymmetric setup in 6 MV photons

Diode out

p

ut (n

ormalized

to

180°)

Gantry angle (degree with 180 being vertical)

Diode (type 473 [12–17 MV rated]) output vs. incident angle;

symmetric setup in 6 MV photons

Diode out

p

ut (n

ormalized

to

180°)