2.3 Electron Detectors

These detectors are used in the measurement of radiation in the range of 5 MeV to 25 MeV. They are
available with a negative or positive output. They are color coded for ease of identification. Following is a
list of the detectors, polarity, color codes, and range.

Model Number

Range

Polarity

Color

30-475

5-25 MeV Electron

Positive

Gray

30-475-8000

5-25 MeV Electron

Negative

Gray


The semiconducting material in these diodes is planar and the diode as a whole is hemispherical in
shape, thus they have a preferred direction. Following are typical graphs of the angular dependence of
these electron detectors.

The output of a typical detector is 1.00 nano coulombs/RAD (1.00 nC/cGy).

Angular Dependence

The electron diodes in a 20 MeV electron beam, shown in the following two figures have a I % change
after a 30 degrees deflection, in both planes from 180.

Gantry angle (degree with 180 being vertical)

Diode (type 475 [electron diode]) output vs. incident angle;

asymmetric setup in 20 MeV electrons

Diode out

p

ut (n

ormalized

to

180°)