Operation

Photon Detectors

2

2-3























5–11 MV Diodes

Similar results are shown in the following figures for 5-11 MV diodes, although the output changes by less
than 1% for a slightly smaller range of angles. That range being 35-40 degrees from 180 for both planes
of rotation.

Gantry angle (degree with 180 being vertical)

Diode (type 471 [1–4 MV rated]) output vs. incident angle;

asymmetric setup cable faces collimator when at 90 degrees

Diode out

p

ut (n

ormalized

to

180°)

Gantry angle (degree with 180 being vertical)

Diode (type 472 [5–11 MV rated]) output vs. incident angle;

symmetric setup in 6 MV phontons

Diode out

p

ut (n

ormalized

to

180°)