Nuclear Associates 30-471
Operators Manual

3-2

























Based on the characteristics of these detectors, they are used for specific applications such as exit dose
measurements, intracavitary measurements and also with radiation therapy. Given the quick response of
these detectors, they are very useful for verifying the patient’s dose delivered during therapy and also as
a safeguard against misadministered radiation treatments.

With various types of integral filters and build-up shields these detectors can measure radiation
exposures as low as 1 MeV and as high as 25 MV. They are also available with a negative or positive
polarity output. The various types of detectors have been color coded to identify them with a specific
energy range. This allows for ease of identification and avoidance of misuse. Following is a list of diodes
and their associated part numbers that are available.

Model Number

Range

Polarity

Color

30-471

1-4 MV Phonton

Positive

Blue

30-471-8000

1-4 MV Phonton

Negative

Blue

30-472 5-11

MV

Phonton

Positive

Yellow

30-472-8000 5-11

MV

Phonton

Negative Yellow

30-473

12-17 MV Phonton

Positive

Red

30-473-8000

12-17 MV Phonton

Negative

Red

30-474

18-25 MV Phonton

Positive

Green

30-474-8000

18-25 MV Phonton

Negative

Green

30-475

5-25 MeV Electron

Positive

Gray

30-475-8000

5-25 MeV Electron

Negative

Gray

Diode (type 474 [18–25 MV phontons])

output farmer chamber vs. SSD

SSD (cm)

Diode out

p

ut/far

mer chamber (normalized to 100 SSD)