91

Immunity test

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

IEC 60601 test level

3 V

rms

150 kHz to 80 MHz

3 V/m
80 MHz to 2.5 GHz

Compliance level

10 V

10 V/m

Electromagnetic environment - guidance

Portable and mobile RF communications equipment should not be used closer to any part
of the bluephase 20i, including cables, than the recommended separation distance cacu-
lated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

d = 0.35 P

d = 0.35 P 80 MHz to 800 MHz
d = 0.70 P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

Guidance and manufacturer's declaration - electromagnetic immunity

The bluephase 20i is intended for use in the electromagnetic environment specified below. The customer or the user of the bluephase 20i should assure
that it is used in such an environment.

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic

environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured field strength in the location in which the bluephase 20i is used
exceeds the applicable RF compliance level above, the bluephase 20i should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the bluephase 20i.

b

Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.

Rated

maximum

output power

of transmitter

W

0.01

0.1

1

10

100

150 kHz to

80 MHz

d = 0.4 P

0.04

0.13

0.40

1.3

4.0

Separation distance according to

frequency of transmitter

m

80 MHz to

800 MHz

d = 0.4 P

0.04

0.13

0.40

1.3

4.0

800 MHz to

2.5 GHz

d = 0.7 P

0.07

0.22

0.70

2.2

7.0

Recommended separation distances between portable and mobile RF communications
equipment and the bluephase 20i

The bluephase 20i is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the bluephase 20i can help prevent electromagnetic interference by maintaining a miminum distance between portable
and mobile RF communications equipment (transmitters) and the bluephase 20i as recommended below,
according to the maximum output power of the communication equipment.

For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency

range applies.

NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects and people.