Chapter 5

Specifications

All specifications are subject to change without notice.
Typical for 25 °C unless otherwise specified.
Specifications in italic
text are guaranteed by design.

Analog input

Table 1. Generic analog input specifications

Parameter

Conditions

Specification

A/D converters

Four dual 24-bit, Sigma-Delta type

Number of channels

8 differential

Input isolation

500 VDC minimum between field wiring and USB
interface

Channel configuration

Thermocouple sensor type

Differential input voltage
range

Thermocouple

±0.080 V

Absolute maximum input
voltage

±C0x through ±C7x relative
to GND

±25 V power on, ±40 V power off.

Input impedance

5 G

Ω, min

Input leakage current

Open thermocouple detect
enabled

105 nA max

Normal mode rejection ratio

f

IN

= 60 Hz

90 dB min

Common mode rejection ratio

f

IN

= 50 Hz/60 Hz

100 dB min

Resolution

24 bits

No missing codes

24 bits

Input coupling

DC

Warm-up time

30 minutes min

Open thermocouple detect

Automatically enabled when the channel pair is
configured for thermocouple sensors.
The maximum open detection time is 3 seconds.

CJC sensor accuracy

15 °C to 35 °C

±0.25 °C typ,
±0.5 °C max

0 °C to 70 °C

–1.0 to +0.5 °C max

Channel configuration

Channel configuration information is stored in the EEPROM of the isolated microcontroller by the firmware
whenever any item is modified. Modification is performed by commands issued over USB from an external
application, and the configuration is made non-volatile through the use of the EEPROM.

Table 2. Channel configuration specifications

Sensor Category

Conditions

Specification

Thermocouple

J, K, S, R, B, E, T, or N
(Note 1)

8 differential channels

Note 1:

The factory default configuration is Type J.