1-12 System Overview

899493

TempScan / MultiScan User's Manual

A CA-47 cable connects the unit with the computer. The TempScan/1100 or MultiScan/1200 end has one
DB9 connector, and the computer end has two connectors – one for a DB9 and one for a DB25. O
ther crossover-type cables can be used if they are wired as shown in the tables. The tables list the
following four connections from the TempScan/1100 or MultiScan/1200 unit:

• To a DB9 connector configured for RS-232
• To a DB25 connector configured for RS-232
• To a DB9 connector configured for RS-422
• To a Mini DIN8 connector configured for RS-422.

TempScan/1100 or MultiScan/1200

To PC Connection (RS-232)

TempScan/1100 or MultiScan/1200

To PC Connection (RS-232)

DB9 Male

Pin & Signal

Cable

Wirin

g

DB9 Female

Pin & Signal

DB9 Male

Pin & Signal

Cable

Wirin

g

DB25 Female

Pin & Signal

2 RxD-

←--

3 TxD-

2 RxD-

←--

2 TxD-

3 TxD- --

→

2 RxD-

3 TxD- --

→

3 RxD-

5 GND

←→

5 GND

5 GND

←→

7 GND

7 RTS+ --

→

8 CTS+

7 RTS+ --

→

5 CTS+

8 CTS+

←--

7 RTS+

8 CTS+

←--

4 RTS+

TempScan/1100 or MultiScan/1200

To Macintosh Connection (RS-422)

TempScan/1100 or MultiScan/1200

To Macintosh Connection (RS-422)

DB9 Male

Pin & Signal

Cable

Wirin

g

DB9 Male

Pin & Signal

DB9 Male

Pin & Signal

Cable

Wirin

g

Mini DIN8 Male

Pin & Signal

1 GND

←→

3 GND

1 GND

←→

4 GND

2 RTS+ --

→

7 CTS+

2 RTS+ --

→

2 CTS+

4 TxD+ --

→

8 RxD+

4 TxD+ --

→

8 RxD+

5 TxD- --

→

9 RxD-

5 TxD- --

→

5 RxD-

6 CTS+

←--

6 RTS+

6 CTS+

←--

1 RTS+

8 RxD+

←--

4 TxD+

8 RxD+

←--

6 TxD+

9 RxD-

←--

5 TxD-

9 RxD-

←--

3 TxD-

The following text describes the various pin connector signals:

• Transmit Data Negative (TxD-): This output pin transmits serial data to an RS-232 or RS-422 device.

The serial data received is sent with the word length, baud rate, stop bits, and parity configured for the
particular port. This signal is low true.

• Transmit Data Positive (TxD+): This output pin transmits serial data to an RS-422 device only.

The pin functions identically to TxD- except that its polarity is inverted. This signal is high true.

• Receive Data Negative (RxD-): This input pin accepts serial data sent by an RS-232 or RS-422 device.

The serial data received is expected to match the word length, baud rate, stop bits, and parity
configuration of the particular port. This signal is low true.

• Receive Data Positive (RxD+): This input pin accepts serial data sent by an RS-422 device only.

It functions identically to RxD- except that its polarity is inverted. This signal is high true.

• Request To Send Positive (RTS+): This output pin is used as a hardware handshake line to prevent an

RS-232 or RS-422 device from transmitting serial data to the TempScan/1100 or MultiScan/1200 unit
when it is not able to accept it. When automatic RTS/CTS handshaking is selected, the unit will assert
(high) the RTS+ signal when greater than 4096 memory locations are available in its internal buffers. If
available memory drops below 4096 bytes, the unit unasserts (low) the RTS+ signal.

• Request To Send Negative (RTS-): This output pin is used as a hardware handshake line with an

RS-422 device only. The pin functions identically to RTS+ except that its polarity is inverted. This