Function 1 - volume, Function 2 - tube diameter, Function 3 - velocity – Watson-Marlow PF6 User Manual

Page 18: Function 4 - acceleration/deceleration, Instruction handbook

Advertising
background image

INSTRUCTION HANDBOOK

PF6

PF6 IH EN 74-105-022 v1.04.doc

Version: 1.04

Page 18 of 30


Function 1 - Volume
Value: ml
Function 1 informs the system of the volume to be filled.

Value

Option

Min

Max

0.01

9999.9

ml. or gram


The entered value must be between 0.01 and 9999.9.


Function 2 - Tube diameter
Value: Inside diameter of the tube in mm.

Drive type

Tube inside diameter in mm

PF6

0.5

0.8

1.2

1.6

3.2

4.8

6.0

8.0



Function 3 - Velocity
Value: Revolutions per minute (rpm).
Velocity range depends on tube size applied.

Range:

Tube Sizes

Max. Velocity

Max. acceleration

0.5 - 0.8 – 1.2 – 1.6

400

100

3.2

400

100

4.8 - 6.0 – 8.0

400

100


The fastest filling will be carried out at the highest velocity setting but the velocity should always be
adjusted to suit the characteristics of the product and to reduce splashing or foaming.


Function 4 - Acceleration/deceleration
Value: An integer number.

This function offers a choice of values between 1 and 100 dependent on the tube size and drive;
1 = slowest, 100 = fastest.

Tube Sizes

Max. Velocity

Max. acceleration

0.5 - 0.8 – 1.2 – 1.6

400

100

3.2

400

100

4.8 - 6.0 – 8.0

400

100

Advertising