Prime function, Program safety checks, Pump size – Energizer MMT-511 User Manual

Page 159: Pump weight, Remote control, Reservoir

Advertising
background image

Pump Specifications 147

Prime function

■

Is used to fill the infusion set tubing prior to use.

■

Primes are not counted in daily totals but are recorded
separately in the prime history.

■

Fixed prime: 0.1-25.0 units (limited by maximum bolus).

■

Manual prime: unlimited. Fill rate is 1 to 5 units/second.

Program safety checks

■

Over 50 independent safety systems are continuously
monitoring all pump operations.

■

Maximum infusion with single fault condition: 0.0074 ml.

Pump size

■

The Paradigm pump is 2 x 3 x .77 inches (.8 inches in battery
area) [5 cm x 7.60 cm x 2 cm.]

Pump weight

■

The pump weighs less than 100 grams.

Remote control

■

Uses radio signals to allow users to program normal boluses
or to suspend/resume their pumps.

Reservoir

■

Volume: up to 176 units of insulin.

■

Low reservoir alerts at approximately 30, 20, and 10 units of
insulin remaining for U100. (20 and 10 units for U50)

■

The user-filled reservoir is made from impact-resistant,
insulin-compatible polypropylene.

■

A clear window in the pump case allows users to see the
remaining amount of insulin.