16. Fill the reservoir with your insulin. Refer to the Instructions

for Use provided with the reservoir.

WARNING: Your pump will not accurately deliver insulin if air

bubbles are present in the reservoir or infusion set. Take
care to remove air bubbles when filling your reservoir
and priming your infusion set.

17. Attach the tubing connector to the top of the filled reservoir as

described in the instructions that came with the reservoir.

18. Once the tubing connector is securely in place, insert the reser-

voir into the top of the pump case.

19. Once the reservoir is

inserted, rotate the
tubing

connector approxi-
mately 1/2 turn
clockwise until
seated. When seated,
the tubing connec-
tor should be
aligned horizontally
as shown in graphic.

20. Make sure the following screen is showing on the pump. Press

ACT if the screen is not visible.

ACT

ESC

tubing
connector

battery cap

MiniMed

MANUAL PRIME

Disconnect from Body

Insert/Lock Reservoir

Hold ACT to Prime