Getting Started 29

4. After the second set of six beeps, the number of units used will

count up on the screen and begin filling the tubing with insu-
lin. The pump will beep every two seconds.

5. Hold

the

ACT button until insulin droplets form at the infu-

sion set tip.

6. When droplets form at the tip of the infusion set and you do

not have any bubbles in your infusion set, release the ACT but-
ton and press ESC. Your Manual Prime is complete. Fixed
Prime will automatically be selected in the Prime Menu for
two minutes.

7. Insert your infusion set per the Instructions for Use provided

with the set.

8. If you have NOT performed steps 4, 5, and 6, do not attach the

infusion set connection to the pump tubing.