Command 3:
Sampling and Trigger Settings

{ 3, Sampling Interval, Number of Samples, Record Time, Trigger Source, Trigger
Threshold, Trigger Edge, Clock Source }

1. Sampling Interval (sec)

0.00002–0.0001: Fast sampling
0.0001–300: Normal sampling
300–16000: Extended sampling

This parameter is ignored when click edge is not timer.

2. Number of Samples

1 to 120000

Limited only by available memory.

–1

Sampled with each data request
Real-time (simultaneous) sampling

3. Record Time

0 None

1 Absolute time recording

2 Relative time recording

4. Trigger Source

0

None
Start sampling immediately upon receipt of Command 3.

1

[Start/Stop]
Start sampling when [Start/Stop] key is pressed.

2

CH1 Trigger

3

CH2 Trigger

4

CH3 Trigger

10 Microphone Trigger

11 SONIC Trigger

Start sampling in accordance with Motion Sensor (EA-2) sampled data.
Trigger monitor interval is in accordance with sampling interval.

12 SONIC Trigger

Same as 2–7.

20 Countdown

When [Start/Stop] key is pressed or when Command 8 is received,
countdown is performed by beeping at 1-second intervals the number of
times specified by count number, and then sampling starts.

–1 Wait for Command 8

When 1–12

Upon receipt of Command 8, sampling starts regardless of specified trigger
state.

5. Trigger Threshold

Hard trigger

(Trigger Source = 2, 3, 4, 12)

Hysteresis

Trigger
threshold

No triggering when hysteresis
part is not passed.

Hysteresis

Ϯ10V (1pin)

Approximately

Ϯ0.27V

Temperature °C

Approximately

Ϯ2°C

Temperature °F

Approximately

Ϯ3.6°F

0 to 5V (Pin 6)

Approximately

Ϯ67.5mV