Energizer PORTABLE GAS DETECTION INSTRUMENT User Manual

Page 18

Advertising
background image

18

6.4 SPAN CALIBRATION MENU FOR TOXIC GAS ONLY
Note: The MINI-SA software requires that an “AUTO ZERO” be performed as
described in Section 6.3 before the instrument will allow a “SPAN CALIBRA-
TION”. Failure to do so will display the following screen when attempting a
“SPAN CALIBRATION” of the instrument:

AUTO ZERO REQUIRED

BEFORE

SPAN CALIBRATION

ACKNOWLEDGE

Depress both pushbuttons simultaneously to acknowledge this message
and “AUTO ZERO” the instrument per Section 6.3 to procede.
The MINI-SA instrument can be calibrated with span gas supplied by SCOTT
INSTRUMENTS or with a calibration gas supply certified to a concentration
within the following ranges:

CO = 25 to 120 ppm
Cl

2

= 5 ppm only Cl

2

ClO

2

= 5 ppm only Cl

2

HCN = 2.0 to 25.0
H

2

S = 10 to 110 ppm

NH

3

= 10 to 40 ppm

NO

2

= 2.0 to 15.0 ppm

PH

3

= 0.5 to 4.5 ppm

SO

2

= 2.0 to 18.0 ppm

See Section 8.0 for SCOTT INSTRUMENTS supplied calibration gas
cylinders.
To perform “SPAN CALIBRATION” of the sensor, depress the “SELECT”
pushbutton when the “SPAN CALIBRATION” screen is displayed as de-
scribed in Section 6.2. The instrument will display the one of the following
screens:

<gas> SENSOR

For Toxic Gases SPAN CALIBRATION

USING <_> ppm <CALGAS>

CHANGE OK