AMMONIA SPECIFICATIONS

Range:

0 to 200 ppm

Response Time:

T90 < 150 seconds

Operating Temperature Range:

-25 to +30

o

C

(-13 to +86

o

F)

Graphic Display Temperature Range:

-10 to +60

o

C

(+14 to +140

o

F)

Operating Humidity:

15 to 90% RH, Non-Condensing

Recommended Storage Temperature:

0 to 20

o

C

(32 to 68

o

F)

Sensor Accuracy:

±3% of Applied Gas

Linearity:

±3% of Full Scale

Battery Life:

About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode

Sensor Cross Sensitivity:

Carbon Monoxide at 300 ppm: 0 ppm
Hydrogen Sulfide, 15 ppm:

≈ 15 ppm

Sulpher Dioxide, 5 ppm:

≈ 4 ppm

Nitric Oxide, 35 ppm:

≈ 7 ppm

Nitrogen Dioxide, 5 ppm: 0 ppm
Chlorine, 1 ppm

≈ -0.5 ppm

Hydrogen, 200 ppm: 0 ppm
Hydrogen Cyanide, 10 ppm:

≈ 0.5 ppm

Hydrogen Chloride, 5 ppm: 0 ppm
Ethylene, 100 ppm: 0 ppm

Consult Factory for additional information.