Calibration cylinders supplied by SCOTT INSTRUMENTS contain 25 PPM
H

2

S or 50 PPM CO. To use another certified concentration, depress the left

pushbutton below the “CHANGE” command. The one of the following screens
will be displayed:

<gas> SENSOR

For Toxic Gases SPAN CALIBRATION

GAS = <_> ppm <CALGAS>

UP DOWN

Factory preset span calibration gas values:

CO = 50 ppm
Cl

2

= 5.0 ppm

ClO

2

= 5.0 ppm Cl

2

HCN = 10.0 ppm
H

2

S = 25 ppm

NH

3

= 25 ppm

NO

2

= 5.0 ppm

PH

3

= 1.0 ppm

SO

2

= 10 ppm

If you select the “UP” pushbutton, for example, one of the following screens
will display with the number value flashing:

<gas> SENSOR

For Toxic Gases SPAN CALIBRATION

GAS = <_> ppm <CALGAS>

UP SET

When the desired value has been reached depress the “SET” pushbutton
and the “CHANGE/OK” screen will appear again. Select the “OK” pushbutton
and one of the following screens will display:

<gas> SENSOR

For Toxic Gases SPAN CALIBRATION

INJECT <_> ppm <CALGAS>

EXIT