Energizer PORTABLE GAS DETECTION INSTRUMENT User Manual

Page 41

Advertising
background image

41

SULPHUR DIOXIDE SPECIFICATION

Range:

0.0 to 99.9 ppm

Response Time:

T90 < 35 seconds

Operating Temperature Range:

-20 to +50

o

C

(-4 to +122

o

F)

Graphic Display Temperature Range:

-10 to +60

o

C

(+14 to +140

o

F)

Operating Humidity:

15 to 90% RH, Non-Condensing

Recommended Storage Temperature:

0 to 20

o

C

(32 to 68

o

F)

Sensor Accuracy:

±3% of Applied Gas

Linearity:

±3% of Full Scale

Battery Life:

About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode

Sensor Cross Sensitivity:

Carbon Monoxide at 300 ppm: < 3 ppm
Hydrogen Sulfide, 15 ppm: 0 ppm
Nitric Oxide, 35 ppm: 0 ppm
Nitrogen Dioxide, 5 ppm:

≈ - 5 ppm

Consult Factory for additional information.