CARBON MONOXIDE SPECIFICATIONS

Range:

0 to 500 ppm (for CO 500)
0 to 999 ppm (for CO 999)

Response Time:

T90 < 25 seconds

Operating Temperature Range:

-20 to +50

o

C

(-4 to +122

o

F)

Graphic Display Temperature Range:

-10 to +60

o

C

(+14 to +140

o

F)

Operating Humidity:

15 to 90% RH, Non-Condensing

Recommended Storage Temperature:

0 to 20

o

C

(32 to 68

o

F)

Sensor Accuracy:

±3% of Applied Gas

Linearity:

±3% of Full Scale

Battery Life:

About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode

Sensor Cross Sensitivity:

Filtered against H

2

S, 15 ppm: < 0.5 ppm

Filtered against SO

2

, 5 ppm

≈ 0 ppm

Nitric Oxide, 35 ppm: < 3 ppm
Nitrogen Dioxide, 5 ppm:

≈ - 1 ppm

Chlorine, 1 ppm

≈ 0 ppm

Hydrogen, 100 ppm: < 40 ppm
Ethylene, 100 ppm: < 50 ppm
Ethanol, 200 ppm:

≈ 0 ppm

Consult Factory for additional information.