The STEL (Short Term Exposure Level) function measures the level of toxic
gas during a fifteen (15) minute interval and alarms when the preset STEL
value in parts per million is exceeded. The alarm levels are as follows:

STEL

TWA

Ammonia

35 ppm

25 ppm

Carbon monoxide

100 ppm

25 ppm

Chlorine

1 ppm

0.5 ppm

Chlorine Dioxide

0.3 ppm

0.1 ppm

Hydrogen Cyanide

4.0 ppm

2.0 ppm

Hydrogen sulfide

15 ppm

10 ppm

Nitrogen Dioxide

5.0 ppm

3.0 ppm

Phosphine

1 ppm

0.3 ppm

Suphur Dioxide

5 ppm

2 ppm

To enter the TWA/STEL Menu, depress the “SELECT” pushbutton as de-
scribed in Section 6.2 when the TWA/STEL Menu screen is displayed. The
instrument will display one of the following screens:

<gas> TWA/STEL MENU

For Toxic Gases

STEL = 0 ppm
TWA = 0 ppm @ 0:05

RESET SHOW

DISABLE TWA/STEL

STEL and TWA values are calculated. STEL averages exposure for the
latest fifteen (15) minutes. TWA shows the eight-hour average exposure.
Both measures start from zero when the instrument is turned on. The TWA
reading accumulates total exposure since turn-on or “RESET” and is calcu-
lated as if the total exposure occurred in a standard eight-hour shift.
The “RESET/DISABLE” pushbuttons allows the user to reset the values and
time on the display to “0”. One of following screens will be displayed when
the “RESET/DISABLE” pushbuttons are depressed:

<gas> TWA/STEL MENU

TIME AVERAGE VALUES

For Toxic Gases

RESET DISABLE