Npt 0 0 2 – Precision Digital PD662 User Manual

Page 19

Advertising
background image

PD662 Loop-Powered Meter

Instruction Manual

19

Calibrating the Meter (Cal)

To scale the meter without a signal source

refer to Scaling the Meter (SCL), page 18.

The meter can be calibrated to display the process in engineering units
by applying the appropriate input signal.
The use of a calibrated signal source is strongly recommended.

1. Press

the

Up arrow button to scroll to the Calibration menu (CAL)

and press Enter.

2. If using linear signal input conditioning, enter the number of

calibration points (2-32).

3. The meter displays in1. Apply a known signal and press Enter.

The display will flash while accepting the signal.

4. When the meter displays d1, press Enter. Enter a corresponding

display value for the signal input, and press Enter to accept.

5. The meter displays in2. Apply a known signal and press Enter.

The display will flash while accepting the signal.

6. When the meter displays d2, press Enter. Enter a corresponding

display value for the signal input, and press Enter to accept.

Recalibrating the Internal Calibration Reference (Ical)

The Internal Calibration (ICAL) menu, located in the Advanced Features
menu, is used to recalibrate the internal calibration reference.
Recalibration is recommended at least every twelve months. Refer to
Internal Calibration (ICal), page 23 for instructions.

d1

Press Enter to
Accept Setting

in2

Calibrate

Input 2

in1

ENTER

Set Display 1

Value

Press Up to
Set Digit Value

Press Right to
Select Next Digit

Press Menu to
Exit at any Time

0

00.0

MENU

npt

0

0

2

Number of Points menu
not displayed for square
root input conditioning.

Set Number of
Scale Points

Cal

Display Flashes

Accepting Input

Advertising