Analysis based on trace, Analysis based on trace -9, O analysis based on trace – RIGOL DS6000 Series User Manual
Page 201

RIGOL
DS6000 User’s Guide
10-9
Analysis Based on Trace
Press Analyze and select “Trace”. Then, set the templet used in analysis based on
trace through the method below.
1. Trace
Press Trace to select the creation method of analysis templet.
Current Frame: select the current frame as the analysis templet.
Average: select the average of the current data frame as the analysis
templet.
2. Setup Template
Press Setup Template to configure the templet immediately. After the analysis
starts, the oscilloscope compares each frame with the templet to compute the
difference and determines whether error frame exists by comparing the
difference with the threshold currently set.
3. TemplateDisp
Press TemplateDisp to enable or disable template display.