3 To Set the Horizontal System
RIGOL
DS6000 User’s Guide
3-9
Horizontal Scale
Being similar to “Vertical Scale”, the horizontal scale adjustment has two modes:
“Coarse” and “Fine”.
Press MENU ScaleAdjust in the horizontal control area (HORIZONTAL) at the
front panel to select the desired mode. Turn HORIZONTAL SCALE to adjust the
horizontal scale. Turn clockwise to reduce the horizontal scale and turn
counterclockwise to increase the scale.
The scale information (e.g.
) at the upper left corner of the screen will
change accordingly during the adjustment. The range of horizontal scale adjustment
is from 500 ps to 1.000 ks.
Coarse (take counterclockwise as an example): set the horizontal scale in 1-2-5
step namely 500 ps, 1 ns, 2 ns, 5 ns…1.000 ks.
Fine: further adjust within smaller range.