Power-on inspection, Power-on inspection -8 – RIGOL DS6000 Series User Manual
Page 30
Advertising

RIGOL
DS6000 User’s Guide
1-8
Power-on Inspection
When the oscilloscope is energized, press the power key
at the lower-left corner
of the front panel to turn on the oscilloscope. During the start-up process, the
oscilloscope performs a series of self-tests and you can hear the sound of relay
switching. After the self-test is finished, the welcome screen is displayed and you can
view the Option Type, Option name, Option Edition and Left time of the option
currently installed in the “Installed Options” pop-up dialog box on the screen. When
the instrument is shipped, a trial version of the option is provided and the left time is
about 2000 minutes. Press UTIL System SelfCalInfo to view the self-test
results.