Troubleshooting
To help you solve commonly encountered problems we have listed some
typical issues with their respective solutions. If the problems persist contact
RIGOL and prepare your device information (System Information
System Info).
1. The screen still dark (no display) after power on:
(1) Check if the fan is running:
—
If running, the connection of the internal screen cable may be
loose.
—
If not, the instrument has failed to power up, please refer to step (2)
or (3).
(2) If the battery (China only) is used for power supply:
—
Check whether the battery is in good condition and whether it is
correctly installed.
—
Check whether the battery status indicator
at the front panel
goes on; if not, the battery contains no electricity and please
charge the battery.
(3) If AC power is used for power supply:
—
Check if the power supply has been connected correctly and power
switch has been turned on.
—
Check if the power fuse is burned. If a new fuse needs to be
installed, please select one with these specifications: 250V AC, T2A;
5 mm×20 mm.
2. Unresponsive keys or wrong response
(1) Press all the keys on the front panel to make sure that each of them is
normal after power on.
(2) Press System Self-Test Key Test to see if all keys are working
properly.
(3) If a key is not working, do not disassemble the instrument by yourself