Specifications – Amprobe HD110C Digital-Multimeter User Manual

Page 12

Advertising
background image

6

Specifications

General Specifications

Display: 3-1/2 digit LCD, 1999 counts, with
annunciators and menu features

Polarity Indication: Automatic

Input overload indication: 1

Low Battery Indication: B < 50 hours battery life
remain, accuracy is no longer guaranteed

Display Update Rate: 2.5/sec, nominal

Oper. Temp. (< 75 % R.H.): 0 °C to +50 °C

Storage Temp: -20 °C to 60 °C, 0 to 80 % RH,
battery removed

Temperature coefficient: 0.1 x spec. accuracy /°C
(0–18 °C and 28–50 °C) all functions and ranges
except for mVDC

Temperature coefficient: 0.3 x spec. accuracy /°C
(0–18 °C and 28–50 °C) for mVDC

Environment: Indoor use only.

Altitude: Up to 2000 m

Power: Standard 9 V battery, NEDA 1604, JIS
006P,IEC 6F22

Auto Power-Down: Meter powers down after
approx. 30 min. of inactivity.

Battery Life: 250 hours typical with carbon-zinc,
500 hours typical with alkaline

Dimensions, with holster (H x W x D): 200 x 102 x
59 mm (7.9 in x 4.0 in x 2.3 in)

Weight (incl. battery): 642 g (1.4 lb)

Accessories: Test leads, battery (in instrument),
hex wrench in holster, Magne-Grip flex strap and
Users Manual

Case material: Flame retardant, high-impact
thermoplastic

Safety: meets EN 61010-1 Cat IV - 1500 V dc and
1000 V ac rms, Class II EN60529:IP67

EMC: Meets EN 61326-1

This product complies with requirements of

the following European Community Directives:
89/336/EEC (Electromagnetic Compatibility) and
73/23/EEC (Low Voltage) as amended by 93/68/EEC
(CE Marking).

However, electrical noise or intense
electromagnetic fields in the vicinity of the
equipment may disturb the measurement circuit.
Measuring instruments will also respond to
unwanted signals that may be present within
the measurement circuit. Users should exercise
care and take appropriate precautions to avoid
misleading results when making measurements in
the presence of electronic interference.

Electrical Specifications

Accuracies at 23 °C ± 5 °C, <75 % RH non-
condensing, guaranteed for one year.

DC Volts

Ranges: 200 mV, 2 V, 20 V, 200 V, 1500 V

Resolution: 0.1 mV

Accuracy: All ranges ±(0.1 % rdg+2 dgts)

Input Impedance: 10 MΩ

OL Protection: 1500 V dc/1000 V ac rms

Transient Protection: 12 kV impulse (1.2 µS/50 µS)
based on EN 61010-1:2001 impulse requirement
for at CAT IV 1000 V product. This product should
not be used in installations where transients
exceed 12 kV.

AC Volts (45 Hz to 2 kHz)

Ranges: 200 mV, 2 V, 20 V, 200 V, 1000 V

Resolution: 0.1 mV

Accuracy:
200 mV to 200 V (45-500 Hz): ±(0.8 % rdg+4 dgts)
1000 V (50-60 Hz): ±(0.8 % rdg+4 dgts),
200 mV to 200 V (500 Hz-2 kHz): ±(1. 5% rdg+5
dgts),
1000 V (60-500 Hz): ±(1.5 % rdg+5 dgts)

Input Impedance: 10 MΩ