Loop – Amprobe DGC-1000A Ground-Resistance-Tester User Manual

Page 11

Advertising
background image

8

HO

LD

OFF

FU

NC

RE

C

R

LOOP

NOTE: If the HI value is set at OL, or the LO value is set at 0, the
ALARM function will be disabled. So they are method to disable one
of the HI or LO alarm.

NOTE:

The HI value can’t be smaller than the low value. And the LO

value can’t be larger than the HI value. HI value will be adjusted to LO
value plus 1 when roll-over occurs. The maximum value of LO value is
HI value minus 1.

NOTE: If data logging is progressing, sound of beeping will be
disabled to save battery power. But the LCD still shows the warning
letters of “HI—“ or “LO—“.

NOTE: The values for the high and low alarm are stored in the
memory. They are restored when the power is turned on.