93 ω, Warning – Amprobe Multitest-1000 Continuity-Tester User Manual
Page 40
7. Press
the
START/STOP key. The instrument starts the test.
)
ρ
36.5
Ω
m
2
0.93 Ω
At the end of the test the instrument
displays the values alongside.
Earth resistivity value
WARNING
Never disconnect the test leads while the word “Measuring” is being
displayed.
SAVE
SAVING:
The tests can be stored by pressing the SAVE key twice
(refer to paragraph 5.1).
Note: If
START/STOP key is pressed & held, the instrument takes more consecutive
tests.
8. When a new value is acquired the instrument emits a short sound and calculates the
new average value.
9. Press
DISP key to show how many measurements are in average value
calculation and the measuring mode.
DISP
10. Press
CLR key if you want to cancel the medium value of the resistance
and the no. of measurements which are included in the calculation.
CLR
Note: Any test with result over 1999
Ω is not inserted in the calculation of medium value.
Example: 1
st
measurement (D=1):
main display: 6,6
Ωm
secondary display: 6,6
Ωm
2
nd
measurement (D=1):