Lcd error messages -42, Rosemount 5300 series, Reference manual – Emerson Process Management ROSEMOUNT 5300 User Manual

Page 172: Lcd error messages

Advertising
background image

Reference Manual

00809-0100-4530, Rev BA

July 2009

Rosemount 5300 Series

7-42

LCD ERROR MESSAGES

Figure 7-24. The 5300 Display
Panel displaying an error
message.

Table 7-11. Error messages
displayed on the 5300 Display
Panel.

For more information on errors, see “Errors” on page 7-36.

3

Error Message

Error Message

Description

RAM FAIL

An error in the gauge data memory (RAM) has been
detected during the startup tests. Note: this resets the
gauge automatically.

FPROM FAIL

An error in the gauge program memory (FPROM) has
been detected during the startup tests. Note: this
resets the gauge automatically.

HREG FAIL

An error in the transmitter configuration memory
(EEPROM) has been detected. The error is either a
checksum error that can be solved by loading the
default database, or a hardware error.
NOTE: the default values are used until the problem is
solved.

OMEM FAIL

MWM FAIL

An error in the microwave module.

DPLY FAIL

An error in the LCD.

MODEM FAIL

Modem hardware failure.

AOUT FAIL

An error in the Analog Out Module.

OHW FAIL

An unspecified hardware error has been detected.

ITEMP FAIL

An error in the internal temperature measurement.

MEAS FAIL

A serious measurement error has been detected.

CONFIG FAIL

At least one configuration parameter is outside the
allowed range.
NOTE: the default values are used until the problem is
solved.

SW FAIL

An error has been detected in the transmitter software.

Advertising