B.2

International Electrotechnical Commission (IEC)
Classification of Hazardous Locations,

Continued

Certification and
Conformity Details,
continued

Table B-4

Standards Australia (LOSC) Certification

Code Description

4H

Intrinsically Safe Ex ia IIC T4 Class I Zone 0.

Flameproof Ex d IIC T6 Class I Zone 1

Non-Sparking Apparatus - Type of Protection ‘n’ Ex n IIC T6
Class I Zone 2

LOSC Intrinsic Safety Parameters

(1)

Ui = 42.4 V

Ii = 225 mA

Pi = 1.2 W

Ci = 4.2 nF

Li = 0

With no integral indicator, or with
integral Smart Meter, option SM.

L

i

= 150

µH

With Analog Meter, option ME.

(1)

Install in accordance with Honeywell drawing 51204309

.


Table B-5

Zone 2 (Europe) Declaration of Conformity

Code Description

3N

Electrical Apparatus With Type of Protection “n” per IEC 79-
15. IP 66/67.

Ex II 3 GD T

(1)

X (Council Directive 94/9/EC)

−40 ≤ Ta ≤ 93°C.

Zone 2 Parameters

U

i

≤ 42 V

I

i

≤ 22 mA

Temp. Code

(1)

T4 at Ta 93

°C Maximum Ambient

Temp. Code

(1)

T5 at Ta 80

°C Maximum Ambient

Temp. Code

(1)

T6 at Ta 65

°C Maximum Ambient