Honeywell ST 3000 User Manual

Page 119

Advertising
background image

10/04

34-ST-99-36 (Addendum to 34-ST-33-39)

5 of 10

Specific Parameters
for Intrinsic Safety

Field wiring terminals, (

+

,

):

Without local analog meter, ME:

With local analog meter, ME:

With local smart digital meter, SM:

Ui = 30 V,

Ii = 100 mA, Pi = 1.2 W

Ci = 4.2 nF, Ri = 0,

Li = 0

Ci = 4.2 nF, Ri = 0,

Li = 150 µH

Ci = 4.2 nF, Ri = 0,

Li = 0

The pressure transmitter is an intrinsically safe apparatus that can be installed in
potentially explosive atmospheres.

The power terminals (+, -) must be connected only to a certified associated
intrinsically safe apparatus.

The electrical parameters (U, I, and P) of the associated apparatus connected to the
power terminals (+, -) must not exceed the following values:

Ui

≤ 30V

Ii

≤ 100 mA

Pi

≤ 1,2 W

Ambient temperature: - 50ºC to 93ºC

NOTE: -50ºC to 93ºC is the certification and “Operative Limits” for the product
family. Refer to individual Specification Sheets for the standard “Rated Condition”
ambient limits for a particular model that, as shown on the data-plate and
certification nameplate, may be less than the certification limits.

Temperature classifications:

IS (ia) 4 – 20 mA / DE

T4 up to Ta

≤ 93ºC

T5 up to Ta

≤ 85ºC

T6 up to Ta

≤ 70ºC

Flameproof (d)

T5 up to Ta

≤ 93ºC

T6 up to Ta

≤ 78ºC

Special conditions
for safe use,

Intrinsic Safety (X)

Enclosure classification: IP 66/67, Type 4X

Specific Parameters
for Flameproof
Installation

Power supply to field wiring terminals, (+, ): Ucc

≤ 42 V

Output Signal: 4–20 mA

Special conditions
for safe use,
Flameproof
Installation

Ambient operating temperature: - 50 to 93ºC

NOTE: -50ºC to 93ºC is the certification and “Operative Limits” for the product family.
Refer to individual Specification Sheets for the standard “Rated Condition” ambient
limits for a particular model that, as shown on the data-plate and certification
nameplate, may be less than the certification limits.