A.4

Setting Range Values (Local Zero and Span),

Continued

Procedure, continued

Table A-3

Setting Range Values Using Local Zero and Span

Adjustments,

Continued

Step Action

3

Loosen end-cap lock and remove end-cap from PWA side of

electronics housing to expose local zero and span assembly or smart

meter with zero and span adjustments.

Example – Local zero and span assembly.

SPAN

ZERO

Example – Smart meter with zero and span adjustments.

%

100

0

UPPER

VALUE

UNITS

LOWER

VALUE

S E T

VAR

SEL.

SPAN

ZERO

Continued on next page