To check the power supply voltages

Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and
assemblies.

W A R N I N G

Hazard voltages exist on the power supply, the CRT, and the CRT driver board. This
procedure is to be performed by service-trained personnel aware of the hazards involved,
such as fire and electrical shock.

1

Turn off the instrument, disconnect the power cord, then remove the top cover.

2

Apply power to the instrument.

3

Using a DVM, measure the power supply voltages.

4

Note problems with the power supply, then return to the power supply flowchart.

Power Supply Voltages

Pin

Supply/Signal

Voltage

Pin

Supply/Signal

Voltage

1-5

5 volt

+5 V

21

Ground sense

0 V

6

+5 volt remote sense

+5 V

22

Power valid

+5 V

7-9

+3.5 volt

+3.5 V

23

3.25 volt remote sense

3.25 V

10

+3.5 volt remote sense

+3.5 V

24-30

3.25 volt

3.25 V

11

+12 volt

+12 V

31

5.2 volt current limit

+2.5 V

12

12 volt

12 V

32

5.2 volt remote sense

5.2 V

13

Not connected

33-37

5.2 volt

5.2 V

14-20

Ground

0 V

38

Remote on/off

+12 V

Troubleshooting
To check the power supply voltages

5–24