To remove and replace optional modules or filler panels

1

To reconfigure your system later in this procedure, note the configuration of your
system.

Some modules for the Logic Analysis System require calibration if you move them to a
different slot.

2

Starting from the top, loosen the thumb screws on filler panels and cards in the
mainframe or the expansion frame. Starting from the top, pull the cards and filler
panels out.

All multicard modules will be cabled together. To prevent damage to the cables and
connectors, pull all multicard modules out together.

3

To replace the modules, reposition all cards and filler panels so that the endplates
overlap properly.

4

Firmly seat the bottom card into the backplane connector of the mainframe. Keep
applying pressure to the center of the card endplate while tightening the thumb
screws finger tight. Repeat for all cards and filler panels in a bottom to top order.

For correct air circulation, filler panels must be installed in all unused card slots. Correct air
circulation keeps the instrument from overheating. Keep any extra filler panels for future use.

See Also

Service Guides for each individual module.

Replacing Assemblies

To remove and replace optional modules or filler panels

6–3