76xx interface board, Pneumatic module, Pneumatic pump – Fluke RUSKA 7350 User Manual

Page 20: Gearmotor, System valve

Advertising
background image

RUSKA 7350

Users Manual

2-4

76xx Interface Board

This board interfaces and conditions the pump position signal. The board interfaces

these signals between the electronic and pneumatic module. This board also provides

a 0 – 10 VDC signal to 0 – 100 psi electronic pneumatic regulators and the motor

controller.

Pneumatic Module

The pneumatic section shown in Figure 2-2 houses components that control and

accurately measure pneumatic pressure.

TRANSDUCER

RELIEF

VALVE

RELIEF

VALVE

ELECTRONIC

REGULATOR

SOLENOID

VALVE

PNEUMATIC

PUMP

VARIABLE

SPEED

GEARMOTOR

AIR-OPERATED

VALVE

TEST

PORT

SUPPLY

PORT

AIR SUPPLY

110 PSI

gld02.eps

Figure 2-2. Pneumatic Diagram

Pneumatic Pump

The pneumatic pump is a positive displacement pump that precisely varies the system

pressure by compressing and expanding system gas. The pump is driven by a variable

speed gear-motor. The pump has a position sensor that tracks the position of the pump

plunger. It has two travel limit switches at each end of its stroke.

Gearmotor

The variable-speed gear motor drives the pneumatic pump. The gear box on the motor

reduces the motor speed.

System Valve

The system valve isolates the pump and system from the gas supply regulators. It is a

high pressure air-operated valve. This valve has a diaphragm operator that requires a

minimum of 60 psi of air pressure to close the valve. The air is supplied to this valve

through a three-way normally-open solenoid valve.