Entering/exiting control mode, Vent, Step/jog – Fluke RUSKA 7350 User Manual

Page 31: Stepping, Setting step size, Jogging

Advertising
background image

Local Operation

Vent

4

4-7

Note

The set-point can also be changed using either the Step or the Jog function.
Please see Chapter 4, Step/Jog.

Entering/Exiting Control Mode

1. The Control mode is set from the Main Menu. (Press PREVIOUS until the Main

Menu appears.)

2. Press CONTROL [F2] to enter Control mode. ENTER must be pressed to confirm

entry into Control mode.

3. Press MEASURE key found in the keypad to exit Control mode. No confirmation is

necessary.

Vent

The Vent [F3] function is used to rapidly vent the pressure in the system to atmosphere.

Step/Jog

In addition to entering the pressure set-point through the key-pad, the pressure set-point

can be changed using either the step or the jog functions. The Step function is used

primarily when the operator is taking pressure steps in equal pressure increments. The

size of the pressure step is user defined. The Jog function is most often used when

calibrating mechanical gauges such as a dial gauge and the operator wishes to change the

pressure until the mechanical gauge indicates a cardinal point. The operator can then read

the higher resolution Calibrator to determine the actual pressure value when the

mechanical gauge is indicating a cardinal point. The Step/Jog [F5] key toggles between

Step and Jog, to switch between these modes, press the function key 5 [F5].

Stepping

Assure that the word Step is highlighted in the Step/Jog [F5]. Using the rotary knob,

rotate the knob clock-wise to increase the set-point, or counter clock-wise to decrease the

controller set-point. The scratchpad will be updated to a new set-point. The step size

defaults to 10% of full scale. Press ENTER to accept the new set-point.

Setting Step Size

From the Main Menu, press Menu [F6]

⎢Setup [F2 ⎢User [F2]. Using the rotary knob,

move the cursor to highlight Step Size. Enter the desired step size in the current pressure

units and press ENTER.

Jogging

Assure that the word Jog is highlighted in the Step/Jog [F5]. From the Main Menu, the

pressure set-point can be modified by rotating the rotary knob clockwise to increase

pressure and counterclockwise to decrease pressure. Each increment or decrement will

change the pressure in the least significant digit. Continually rotating the rotary knob, the

pressure will continue to change until the rotation is stopped.