Date/Time

The Calibrator’s system clock is continuously updated. If the date or time requires editing,

the following applies;
1. The date and time are set from the Setup | System | Menu. From the Main Menu

(press PREVIOUS until the Main Menu appears), press Menu [F6] | Setup [F2] |

System [F5].

2. To set the system date, turn the rotary knob until Date is highlighted. Use the numeric

keypad to enter the current four-digit year, month, and day (yyyymmdd). All digits

must be entered. Press ENTER to accept.

3. To set the system time, turn the rotary knob until Time is highlighted. Use the

numeric keypad to enter the current hour, minute, and second (hhmmss). All digits

must be entered. Press ENTER to accept.

Reset

The Reset [F1] command is used to re-boot the Calibrator.
To reset the system, from the Main Menu (press PREVIOUS until the Main Menu

appears), press Menu [F6] | Setup [F2] | System [F5] | Reset [F1].

Menu | Calibrate

The Menu | Calibrate command is used to perform the calibrations on all of the sensors

utilized by the Calibrator. The RUSKA 7350 can be completely calibrated using the front

panel interface. It does not require any external computer or software in order to perform

a successful calibration. It also has the capability of being calibrated remotely through

either the RS232 or IEEE-488 interface. The RUSKA 7350 will guide the operator

through a menu driven calibration procedure identifying the pressures that should be

supplied by a standard. Following the calibration procedure, the RUSKA 7350 calculates

and saves its own coefficients. The calibration coefficients are date stamped so the user

can easily identify the last time the instrument went through a full calibration, when the

coefficients were last edited or when the Calibrator was re-zeroed.
The calibration section in the RUSKA 7350 can be password protected to prevent any

unauthorized entry into the calibration section of the system. Please refer to Chapter 6,

Maintenance, of the manual for more detailed information regarding the calibration of the

RUSKA 7350.

Calibration Password

The calibration password allows the user to protect access to Calibrator calibration

constants and the calibration procedure. If the calibration password is set to any number

other than zero, it is required before the user is allowed to calibrate the Calibrator or

manually change the calibration constants.

Note

It is recommended that the access password be recorded and filed in a
secure location.

1. The calibration password is set from the Calibration Menu. From the Main Menu

(press PREVIOUS until the Main Menu appears), press Menu [F6] | Calibrate [F3] |

Menu | Calibrate | Edit.

2. Press the Access [F4] key. If there is a password in the system other than 0, then you

must enter the current valid password into the RUSKA 7350 before it will allow you

to modify the password.