Ruska 7350 pump lubrication, Approved lubrications, Frequency of lubrication – Fluke RUSKA 7350 User Manual

Page 68: Lubrication procedure, Fuse, Cleaning

Advertising
background image

RUSKA 7350

Users Manual

6-6

RUSKA 7350 Pump Lubrication

W

Caution

The threaded portion of the RUSKA 7350 pump spindle requires

periodic lubrication. Failure to lubricate the pump spindle can

cause premature failure of the pump spindle and nut resulting

in permanent pump damage.

Approved Lubrications

Jet Lube AP-5 Grease (part number 45-342)
Jet Lube MP-50 Moly-paste

Frequency of Lubrication

The frequency of lubrication is dependant upon the frequency of instrument use. On a

highly utilized system, the pump must be lubricated every other week. Also the pump

should be lubricated any time that the pump shows signs of loss of lubrication along the

face of the pump threads.
If a pump position error occurs, along with checking the condition of the pump and pump

over-travel limit switches, please check the pump lubrication. An inadequately lubricated

pump can sometimes be the cause of pump-position error detection by the system

software.

Lubrication Procedure

The threaded portion of the pump plunger requires periodic lubrication. The pump may

be lubricated by either injecting the lubricant through the grease fitting located on the left

side of the instrument or by applying the grease directly to the threads of the pump

plunger. Either way, the power to the system should be removed when applying the

lubrication to the pump.
When applying grease directly to the pump, extend the pump plunger such that the

majority of the threads are exposed. Using the lubrication provided with the system, or

one of the lubrications noted in the approved lubrication list, disperse a generous amount

of lubrication all along the threaded portion of the pump plunger.

W

Caution

Only the lubrication defined on the approved lubrication list

above can be used. Substituting this lubrication can cause

permanent damage to the pumping mechanism.

Fuse

The main fuse is a 6 AMP, 250V type 3AB-Slow blow (part number 3926870).

Cleaning

When necessary, clean externally using a damp lint-free cloth and mild liquid detergent.

Advertising