Start-up error messages, Downloading recorded data, Placement of ir dongle for transfer of data – Fluke 5020A User Manual

Page 135

Advertising
background image

Maintenance and Troubleshooting

Downloading Recorded Data

8

8-5

SENSOR <n>: CALIBRATION INVALID

Cycle the power. Try a different sensor. Contact an
Authorized Service Center.

SENSOR <n>: CALIBRATION EXPIRED

Contact an Authorized Service Center to have the
sensor recalibrated. Check the time/date to make
sure it is correct.

POWER DOWN: CALCULATION ERROR

Contact an Authorized Service Center.

POWER DOWN: SOFTWARE RESET

Contact an Authorized Service Center.

POWER DOWN: HARDWARE RESET

Contact an Authorized Service Center.

BUTTONS: <button name>...

Check to make sure the buttons are not being
pressed. Cycle the power. Contact an Authorized
Service Center.

Start-up Error Messages

Error messages may appear after the self-test. If this happens, contact a Fluke Authorized

Service Center for assistance.

Downloading Recorded Data

The thermo-hygrometer stores up to 400,000 readings in its memory. These readings can

be downloaded to a computer using the RS-232 serial port.
The following steps should be followed to download logged data using the serial port:

1. Connect the thermo-hygrometer to a COM port on the computer using the serial

cable.

2. Run a terminal program such as Windows(r) Terminal or HyperTerminal(r).

a. Set up the terminal program to use the correct COM port and baud rate.

Configure the communication settings to use 8 data bits, 1 stop bit, no parity

and no flow control.

b. If you want to save the downloaded data to a file, configure your terminal

program to capture the incoming text and save it to a text file. Refer to your

terminal program’s documentation for details on how to do this.

3. Begin downloading the data.

a. Use the menu function DATA RECORD - DATA PRINT to begin sending

recorded data uning text format. See Data Print.

b. The thermo-hygrometer should immediately begin sending the data. If no

readings have been stored, the thermo-hygrometer will not respond to these

commands.

4. If the terminal program was set to capture the incoming data to a file, wait until

no more data is being received, then stop capturing.

The recorded data can also be transferred using the IR port with LogWare III software, or

written to a file with a data card.

Placement of IR Dongle for Transfer of Data

The thermo-hygrometer can be used with the LogWare III software to collect data via the

IR interface. Place your device near the IR port of the thermo-hygrometer between 3 and

6 inches (8 and 16 cm) and pointed directly at and perpendicular to the IR window.