Digital Communications Interface

Commands

6

6-15

Response: 0
A value of 1 is returned if a power loss alarm event has occurred. A value of 0 is returned

if a power loss alarm event has not occurred.

ALARm:POWer:ENABle?

Returns the state of the power loss alarm enable.
Example: ALAR:POW:ENAB?
Response: 0
A value of 1 is returned if the power loss alarm is enabled. A value of 0 is returned if the

power loss alarm is disabled.

ALARm:POWer:ENABle <bool>

Sets the state of the power loss alarm enable.
Example: ALAR:POW:ENAB 1
The <bool> parameter turns the power loss alarm on (1 or ON) or off (0 or OFF). This

command may be password protected (see Password Commands).

ALARm:RHUMidity<chn>:LOWer?

Indicates whether a lower humidity alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:LOW?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if a

lower humidity alarm event has occurred on the specified channel. A value of 0 is

returned if a lower humidity alarm event has not occurred on the specified channel.

ALARm:RHUMidity<chn>:LOWer:ENABle?

Returns the state of the lower humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:LOW:ENAB?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if

the lower humidity alarm is enabled. A value of 0 is returned if the lower humidity alarm

is disabled.

ALARm:RHUMidity<chn>:LOWer:ENABle <bool>

Sets the state of the lower humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:LOW:ENAB 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <bool> parameter

turns the lower humidity alarm enable on (1 or ON) or off (0 or OFF). This command

may be password protected (see Password Commands).

ALARm:RHUMidity<chn>:LOWer:LIMit? [MIN|MAX|DEF]

Returns the lower humidity alarm limit for the specified channel in %RH.
Example: ALAR:RHUM1:LOW:LIM?