Caution

When data recording memory fills up, the earliest block of data

is automatically discarded so recording of the most recent data

can continue.

Data Record

The DATA RECORD function allows the setting of the data recording features. The

functions that appear in this menu are RECORD SETTING, DATA VIEW, DATA

PRINT, DATA STORAGE, and DATA CLEAR.

avu018.jpg

Figure 5-17. Data Record

Use the A D buttons to select the menu. Press the E button to select the menu. Press

the F button to cancel any changes and exit to the menu.

Record Setting

The RECORD SETTING function allows you to select what is to be recorded and how

often the recording occurs. This may be password-protected. The temperature and

humidity from each channel can be recorded and the recording period is selectable.
To record the temperature on sensor channel 1, set REC 1/T to ON.
To record the humidity on sensor channel 1, set REC 1/H to ON.
To record the temperature on sensor channel 2, set REC 2/T to ON.
To record the humidity on sensor channel 2, set REC 2/H to ON.
The REC PER recording period is set by selecting 1, 2, 5, 10, 15, or 30 seconds, 1, 2, 5,

10, 20, or 30 minutes, or 1 hour for the REC PER.

Note

It is recommended that REC PER be set to 5 minutes or more; otherwise, the memory may fill up
quickly and redrawing graphs may be slow. Also note that graphs can only be redrawn if the
measurement for the channel is being recorded, as the graph data is obtained from recorded
data. The recording period affects the observable time resolution of graphs

.