Display layout, Field data – Fluke 5020A User Manual

Page 45

Advertising
background image

Menu Functions

Display Menu

5

5-9

Display Layout

The DISPLAY LAYOUT function allows you configure any of the 16 different display

layouts. These settings may be password-protected. A display layout can be enabled or

disabled by choosing ON or OFF. ON allows the display to be selected from the main

screen using the B C buttons. The number of zones is selectable from 1 to 4. Each zone

can display one of the following zone types: blank screen, time, stats, temperature and

humidity numerically, or temperature and humidity graphically. Once the number of

zones is selected, you can select what is to be displayed in each zone.
The B C buttons are used to select the parameter setting. The A D buttons are used to

move between each parameter. Press the E button to change and save the new settings.

Press the F button to cancel any changes and exit to the menu.
The enabled layouts can be viewed by pressing the B C buttons when the main screen is

displayed.

avu012.jpg

Figure 5-11. Display Layout

Field Data

The FIELD DATA function allows you to select a type of statistic for any of the fields in

a STAT type zone. This may be password-protected. Select the display layout on which

to adjust the statistics field type and then select the zone to adjust. This function can only

be used with display layouts that contain STATS zones. There are eight fields per zone.

Any field can be set to the following field types: DATE, TIME, ID, SERN, TEMP, T

RATE, T AVE, T STD, T MIN, T MAX, T SPR, T N, T RMAX, T ALAR, T DELT,

RHUM, H RATE, H AVE, H STD, H MIN, H MAX, H SPR, H N, H RMAX, H ALAR,

H DELT, DEWP, HEAT, REC CH, BATT, MEAS, or BLANK. See Table 5-1 for a

definition of each of these parameters.
Statistics fields show running statistics of the measurements taken since the last statistics

reset. The statistics are reset automatically after power up, when the STATS RESET

function is used or when the time reaches the RESTIM hour and AUT RES is enabled for

the DAILY STATS.

Advertising