Ce comments, Emc directive, Immunity testing – Fluke 5020A User Manual

Page 136: Emission testing, Low voltage directive (safety), Frequently asked questions

Advertising
background image

5020A

Users Manual

8-6

Note

When using IRCOMM and an IR dongle, the data transfer can be

susceptible to interference from fluorescent lighting.

CE Comments

EMC Directive

Fluke’s equipment has been tested to meet the European Electromagnetic Compatibility

Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your

instrument lists the specific standards to which the unit was tested.
The instrument was designed specifically as a test and measuring device.
Compliance to the EMC directive is through IEC 61326-1 Electrical equipment for

measurement, control and laboratory use - EMC requirements (1998).
As noted in the IEC 61326-1, the instrument can have varying configurations.
The instrument was tested in a typical configuration with shielded, grounded probe and

RS-232 cables. Emissions may, in non-typical applications, exceed the levels required by

the standard. It is not practical to test all configurations, as the manufacturer has no

control over the probes the user may connect to the instrument.

Immunity Testing

The instrument was tested to the requirements for industrial locations.
This allows the instrument to be used in all types of locations from the laboratory to the

factory floor. Criterion B was used for Radiated RF (IEC 61000-4-3) and Conducted RF

(IEC 61000-4-6). Therefore, the operation of the instrument may be affected by excessive

electromagnetic interference and the instrument may not perform within the normal

specification limits in such an environment. Criterion C was used for Electrostatic

Discharge
(ESD, IEC 61000-4-2) and Electric Fast Transit (EFT, Burst, IEC 61000-4-4). If the

instrument is subjected to EFT conditions at 2kV, the instrument may require the user to

cycle the power to return to normal operation.

Emission Testing

The instrument fulfills the limit requirements for Class A equipment but does not fulfill

the limit requirements for Class B equipment. The instrument was not designed to be

used in domestic establishments.

Low Voltage Directive (Safety)

In order to comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and

the IEC 1010-2-010 (EN 61010-2-010) standards.

Frequently Asked Questions

Q. Can the 5020A operate on battery power?
A. No. The 5020A requires AC power to be fully functional. However, it does utilize a

9V backup battery that runs the data collection in the event of a power outage. When AC

power is interrupted, the display will go blank and an audible alarm will sound to alert

you that the 5020A is running on backup power. The 5020A will continue to collect data