Symmetricom 58502A User Manual

Page 25

Advertising
background image

Chapter 1 Getting Started

Using Commands to Control Key Functions (Examples)

User’s Guide

1-15

To Check Instrument Identification

Enter *IDN?, and press Return.

A quoted identification string is returned, similar to the following:

HEWLETT-PACKARD, 58502A,3426A–00123,3422 – A

To Clear the Error Queue

Enter *CLS, and press Return.

The error message disappears and the scpi> prompt returns.

Changing the RS-232 Serial Port Characteristics

Refer to the subsections “Making Changes to the Serial Port Settings
(If Needed)”
and “Determining the Serial Port Settings” on page 3-8 in
Chapter 3 for complete details and cautions concerning the use of these
commands. Always remember to make corresponding changes to your
PC serial-port parameters after changing the instrument’s serial-port
characteristics.

At first power-on from the factory, serial-port settings are as follows:

Full Duplex is ON (can be turned OFF)

Baud rate is 9600 (can also be set to 1200, 2400, or 19200)

Data bits is 8 (set by parity = None)

Pace is NONE (can be changed to XON)

Parity is NONE (can also be set to ODD or EVEN for 7 data bits)

If you need to change one of these, enter the desired parameters
appended to the following commands for the stated purpose.

To Turn Off Full-Duplex Communication

Enter SYST:COMM:SER:FDUP 0, and press Return.

To Turn On Full-Duplex Communication

Enter SYST:COMM:SER:FDUP 1, and press Return.

To Check If Full-Duplex Communication is On or Off

Enter SYST:COMM:SER:FDUP?, and press Return.

If 1 is returned, duplex is on. If 0 is returned, duplex is off.