Test equipment required – Symmetricom 58502A User Manual

Page 30

Advertising
background image

Chapter 2 Performance Tests

Test Equipment Required

2-4

User’s Guide

Test Equipment Required

Table 2-1. Recommended Test Equipment

Instrument

Required Characteristics

Recommended Model

Use

*

* C= Calibration (Option 010 only)

OV = Operational Verification
P = Performance Tests

Signal Generator

Frequency Range: 100 kHz to 10 MHz
Output Level: +6 to +16 dBm
Harmonic Distortion:<-30 dBc
Spurious Distortion:<-80 dBc

HP 8663A (or equivalent)

OV, P

Digitizing
Oscilloscope

Frequency range: 100 kHz to 100 MHz HP 54600B (or equivalent) OV

Universal Counter

Frequency Range: 100 kHz to 100
MHz: >10 digits resolution @ 10 MHz

HP 53131A (or equivalent) C (for

Option 010
only)

10 MHz Reference
Source

10 MHz Reference Signal
Accuracy:

±

2 X 10

-12

HP 5071A (or equivalent)

C

Terminal or computer

Communication software, RS-232
connection

Any Model

OV, P, C

RMS Voltmeter

Accuracy:

±

2% at 3V Full Scale,

1 MHz

HP 3400B (or equivalent)

P

50

Ω BNC

Feedthrough

_____________________________

HP 10100C (or equivalent) OV, P, C

RS-232 Interface
Cable

DTE-to-DTE 9-pin (f) to 9-pin (f)

HP 24542U (or equivalent) C

Type N-to-BNC
Connector

_____________________________

H P 1250-0780 (or
equivalent)

OV, P

50

Ω BNC Coaxial

Cable (2)

RG-58, 12 meter, male-to-male

HP 10503A (or equivalent) OV, P, C