Chapter 2 Performance Tests

58502A Operational Verification

2-6

User’s Guide

3

Connect the equipment as shown in Figure 2-1. (Connect the Signal
Generator output to the 58502A’s input A or B rear-panel input, and
connect oscilloscope Channel 1 to output 1 port of the 58502A.

Figure 2-1. Operational Verification Setup

4

Verify that the front-panel Input A or B and all twelve OUTPUT
STATUS (1-12)
output LEDs are illuminated.

5

Set the oscilloscope to the following:

Channel 1: 0.5 V/div
Timebase: 50 ns/div

6

Connect the oscilloscope to each of the rear-panel output ports in turn
to verify that the output signal is a nominal 1 V rms (2.8 Vp-p) sine
wave at 10 MHz.

7

Disconnect the signal from the 58502A rear-panel input connector.

8

Verify that all input and output LEDs are off.

This indicates loss of input and output.

9

Mark Pass or Fail in Line 1 on the Operational Verification portion of
the 58502A Performance Test Record, located on page 2-14.